THV-10MDX Advanced Automatic Turret Digital Vickers Hardness Tester
Product description:
Product introduction: Using electronic electric loading, high precision sensor, with a unique closed loop control system, every gear force is automatically compensated. Using touch screen interface, menu structure simple operation. With dig
Using electronic electric loading, high precision sensor, with a unique closed loop control system, every gear force is automatically compensated.
Using touch screen interface, menu structure simple operation.
With digital encoder eyepiece structure, measure D1, D2 value, LCD directly display hardness value and D1, D2 value.
Automatic turret, testing head and objective lens automatic switch to each other, test point automatic positioning accurate, automatic loading, automatic unloading.
It can convert Brinell, Rockwell and Vickers hardness values.
It can be calibrated according to standard hardness block or length scale.
The test force can be automatically corrected by standard dynamometer.
All data is saved on USB flash drive in Excel format.
Dual language setting, convenient for users to choose.
Technical parameters:
Model | THV-10MDX |
Test force | 0.3Kgf (2.94N), 0.5Kgf (4.90N), 1.0 Kgf(9.8N), 2.0Kgf(19.6N), 2.5Kgf(24.5N), 3.0Kgf(29.4N), 5.0Kgf(49.0N), 10.0Kgf(98.0N) |
Meet standard | GB/T4340, ASTM E92 |
Minimum unit of test | 0.01 μm |
Data output | LCD display readout, save to U disk in Excel format directly |
Conversion scale | HRA, HRB, HRC, HRD, HRF, HV, HK, HBW, HR15N, HR30N, HR45N, HR15T, HR30T, HR45T |
Hardness test range | 8~2900HV |
Test force application method | Automatic (loading, holding, unloading) |
Test the magnification of microscope | 100X(test), 200X(test) |
Load holding time of test force | 1~99s |
Maximum permissible specimen height | 180mm |
Distance from center of indenter to outer wall | 160mm |
Size | 580*190*700mm |
The weight of the | About 55 kg |
The power supply | AC220V ± 5%, 50 to 60 hz |