THR-45XC Automatic Rockwell hardness (convex nasal)
Product description:
1.Features 1. Specially designed head horizontally protruding ring can be measured, such as general hardness tubular parts can not be measured, the minimum inner diameter of 15mm test 2. Automatic fool type operation, only a key to get resu
1. Specially designed head horizontally protruding ring can be measured, such as general hardness tubular parts can not be measured, the minimum inner diameter of 15mm test
2. Automatic "fool" type operation, only a key to get results without artificial increase preload to avoid human error
3. The touch-screen interface, easy to operate, in any language version
4. There is no need to install weights or installation.
5. Electric load and closed-loop control
6. Test force automatic correction, improve the accuracy of a force of magnitude
7. All Rockwell scale, high, medium and low hardness value is automatically corrected three sections
8. Set up a password-protected setup parameters; keeping more samples and test information
9. U disk save measurement data to EXCEL format for easy editing and processing
10. modular design, easy maintenance
Application Fields:
Industrial production, Inspection of product quality on processing technology. Ferrous and nonferrous metals, Hardened steel, tempered steel, annealed steel, hardened steel, sheet of various thicknesses, carbide materials, powder metallurgy materials, hardness and thermal spray coating.
Sheet metal, thin-walled pipe, hardened steel, and small parts of hardness.
Widely used in machinery manufacturing, metallurgy plants, universities, laboratories, etc.
Main Parameters:
Model | THR-45XC |
Conversion Scale | Rockwell, Brinell, Vickers |
Display of Hardness Value | Touchable Large LCD screen display |
Data output | Inside Printer, RS-232 |
Duration time | 0-60s |
Carried Standard | GB/T230.2 Chinese Standard, JJG112 Inspection Rule |
Preliminary Testing force | 3kgf(29.4N) |
Testing force | 15kgf(147.1N) 30kgf(294.2N) 45kgf(441.3N) |
Hardness measuring range |
HR15N:70-94 HR30N:42-86 HR45N:20-77 HR15T:67-93 HR30T:29-82 HR45T:10-72 |
Test force Deviation | ≤0.5% |
Testing accuracy | 0.1 |
Max Height of Specimen | 300mm |
Distance of Indenter to outer wall | 200mm |
Power supply | AC220V+5%,50-60Hz |
Overall Dimension (mm) | 710mm*250mm*880mm |
Data Output | USB Flash Disk |