CN
 Logo

COMPANY PRODUCT

HRS-150/45DGT Nosed Digital Rockwell Hardness Tester

Product description:

Main features : 1. The specially designed indenter is extended horizontally, which can measure the parts that are unable to be measured by the general hardness tester, such as ring and tube, with the minimum test inner diameter of 25mm。 2

Share With:

We're here to help:
Easy ways to get the answers you need.
Send E-mail Chat Now


Main features:
1. The specially designed indenter is extended horizontally, which can measure the parts that are unable to be measured by the general hardness tester, such as ring and tube, with the minimum test inner diameter of 25mm。
2. Adopting electronic and electric loading, high precision sensor, with unique closed loop control system.
3. Touch screen interface, easy to operate, directly select the hardness scale, the force value is automatically changed in the test parameters.
4. According to the standard hardness block, the three segments of each hardness scale are automatically corrected.
5. The test force can be modified automatically by standard dynamometer.
6. Auto-loading main test force
7. The data processing function is powerful, including: HR, HB, HV and other hardness type conversion, setting up the qualified range, automatic alarm for over-limit, and inputting the information such as tester, sample name, etc.
8. More samples and test information, the measurement data U disk saved in EXCEL format, easy to edit and process.

Technical Parameters:

Model HRS-150/45DGT
Rockwell initial test force 10kgf(98.07N)
Superficial Rockwell initial test force 3.0kgf(29.4N)
Rockwell total test force 60kgf(588N)   100kgf(980N)    150kgf(1471N)
Superficial Rockwell total test force 15kgf(147.1N)   30kgf(249.2N)    45kgf(441.3N)
Rockwell hardness scale HRA、HRB、HRC、HRD、HRE、HRF、HRG、HRH、HRK、 HRL、HRM、HRR、HRP、HRS、HRV   
Superficial Rockwell hardness scale 15N、15T、15W、15X、15Y、30N、30T、30W、30X、30Y、45N、45T 、45W、45X、45Y
Rockwell measurement range HRA:20-88、HRB:20-100 、HRC:20-70 、HRD:40-77 、HRE:70-100 、HRF:60-100 、HRG:30-94、HRH:80-100、HRK:40-100、HRL:100-120  HRM:85-115、 HRR:114-125
Superficial Rockwell measurement range HR15N:70-94 HR30N:42-86 HR45N:20-77
HR15T:67-93 HR30T:29-82 HR45T:1-72
Dwell time 0-99s
Hardness resolution 0.1HR
Force error <1.0%
Data output LCD display
Data storage The measured values are stored in the U disk in EXCEL format
Conversion scale Brinell, Vickers
Carried standard BSEN 6508, ISO 6508, ASTM E18, GB/T230
Maximum permissible specimen height 300mm
Distance from indenter to wall 250mm
Dimension 720×260×840mm
Weight 80kg
Power AC220+5%,50~60Hz