THR-150/45XC Automatic Rockwell hardness (convex nasal)

Product description:

Advantages: 1. Specially designed head horizontally protruding ring can be measured, such as general hardness tubular parts can not be measured, the minimum inner diameter of 15mm test 2. Automatic fool type operation, only a key to get res

Share With:

We're here to help:
Easy ways to get the answers you need.
Send E-mail Chat Now




-THR-150/45XC Automatic Rockwell hardness (convex nasal)
Advantages:
1. Specially designed head horizontally protruding ring can be measured, such as general hardness tubular parts can not be measured, the minimum inner diameter of 15mm test
2. Automatic "fool" type operation, only a key to get results without artificial increase preload to avoid human error
3. The touch-screen interface, easy to operate, in any language version
4. There is no need to install weights or installation.
5. Electric load and closed-loop control
6. Test force automatic correction, improve the accuracy of a force of magnitude
7. All Rockwell scale, high, medium and low hardness value is automatically corrected three sections
8. Set up a password-protected setup parameters; keeping more samples and test information
9. U disk save measurement data to EXCEL format for easy editing and processing
10. modular design, easy maintenance
Application Fields:
Industrial production, Inspection of product quality on processing technology. Ferrous and nonferrous metals, Hardened steel, tempered steel, annealed steel, hardened steel, sheet of various thicknesses, carbide materials, powder metallurgy materials, hardness and thermal spray coating.
Sheet metal, thin-walled pipe, hardened steel, and small parts of hardness.
Widely used in machinery manufacturing, metallurgy plants, universities, laboratories, etc.

Main Parameters:                               

Model Name      THR-150/45XC
Exchanging scales Rockwell, Brinell, Vickers
Dwell time 0-60s
Display of Hardness Value Touchable Large LCD screen display
Data output Built-in Printer, RS-232 
Initial load of Rockwell 10kgf (98N)
Total load of Rockwell 60kgf (588.4N) 100kgf (980.7N) 150kgf (1471N)
Initial load of Superficial Rockwell 3kgf (29.4N)
Total load of Superficial Rockwell 15kgf (147.1N) 30kgf (294.2N) 45kgf (441.3N)
Load value error Less than 0.5%
Hardness testing range HRA: 20-96 HRB: 20-100 HRC: 20-70 HRD: 40-77 HRE: 70-100 HRF: 60-100 HRG: 30-94
HRH: 80-100  HRK: 40-100  HRL: 50-115  HRM: 50-115 HRR: 50-115 
Testing rage of Superficial Rockwell HR15N: 70-94; HR30N: 42-86; HR45N: 20-77;
HR15T: 67-93; HR30T: 29-82; HR45T: 10-72
Hardness measurement accuracy 0.1
Executive standard GB/T230.1  GB/T230.2 national standard  JJG112 verification regulation
Hardness reading LCD display
Data output USB flash disk
Power supply AC220V
Max Height of Specimen 300mm
Distance of Indenter to outer wall 200mm
Net weight About 70kg