CN
 Logo

COMPANY PRODUCT

HBRVS-250DU Universal Hardness Tester

Product description:

Main features : 1.There are three kinds of test methods: Brinell, Rockwell and Vickers. 2.Electric loading, high precision sensor, with a unique closed-loop control system 3. Eyepiece structure with digital encoder, measure D1, D2 value,

Share With:

We're here to help:
Easy ways to get the answers you need.
Send E-mail Chat Now


Main features
1. There are three kinds of test methods: Brinell, Rockwell and Vickers.
2. Electric loading, high precision sensor, with a unique closed-loop control system
3. Eyepiece structure with digital encoder, measure D1, D2 value, LCD display hardness value and D1, D2 value directly.
4. The test force is automatically corrected, each force is automatically compensated, and the accuracy of the force is improved by an order of magnitude.
5. Can be calibrated according to the standard hardness block or length scale.
6. Test force can be automatically corrected by standard dynamometer
7. The hardness value of rockwell and vickers can be converted.
8. Touch screen, easy to operate.
9. According to the national standard /ASTM hardness automatic conversion.
10. Set password protection setting parameters, more sample and test information.
11. The measurement data U disk is saved in EXCEL format for easy editing and processing.
12. You can choose to upgrade and add two rulers, HVT and HBT.
 
 
Technical specification

Model THUS-250
Hardness scale Brinell, Vickers, Rockwell
Test force 5~250kgf
Brinell scale HBW1/:5、10、30kgf
HBW2.5/:6.25、15.625、31.25、62.5、187.5kgf
HBW5/:25、62.5、125、250kgf
HBW10/:100、250kgf
Rockwell scale A, B, C, D, E, F, G, H, K, L, M, P, R, S, V
Vickers scale HV5、HV10、HV20、HV30、HV50、HV100、HV120kgf
Testing system The indentation image is collected by optical objective lens and CCD camera and measured in LCD interface:
Resolution:
The length of reading:0.1μm
    The depth of reading:0.2 μm
Objective magnification 2.5X、  5X( Optional 10X、20X)
Camera 1280*1080 pixel, 1/2 inch
Data display Computer display data
Testing Automatic loading, dwell, unloading
Indenters Diamond Rockwell indeneter: 120°
Rockwell ball indenter: 1/16",
Diamond Vickers indenter: 136°
Brinell ball indenter: 1mm, 2.5mm, 5mm, 10mm
(Rockwell ball indenter options: 1/8", 1/4", 1/2")
Lighting system LED light ,ring light
Standard conforms to GB/T230、GB/T231、GB/T4340、JJG144-1999、GB/T18449、ISO 6508、 ASTM E 10、 ASTM E92、 ASTM E18、 ASTM E384、 ASTM E103
Total test force dwell time 1 - 99 秒
Sample specification Max height of sample: 220mm
Indenetr-outer wall(throat):200mm
Cylindrical specimen:the minimum diameter of cylindrical specimens on the outer surface is 4mm;
Max height of specimen Rockwell:220、Brinell   Vickers:150mm
Distance from center of head to outer wall 200mm
Dimension 560*350*800mm
Weight About 70Kg
Power supply AC220V+5%,50-60Hz